Opis izdelka
It covers a range from 10 °C above ambient temperature up to375 °C. It has a stability of ±0,15 °C and anaccuracy of ±0,25 °C at 150 °C, of ±0,35°C at 375 °C.
The time taken to reach 375 °C is just 20 minutes and ittakes 40 minutes to reach 100 °C from 375 °C. This allowsto save time during calibration operations that requires more thanone verification point. The calibrator has a hole with a diameterof 26x150 mm in which the reduction inserts are inserted to adaptthe diameter of the probe to be calibrated to the calibrator.
There are 3 inserts available with holes ranging from a diameterof 3,2 to 12,7 mm while other holes are provided upon request up to1 mm in diameter.
The appliance's innovative ventilation system allows thecalibrator to keep the temperature on the top of the calibratorlower compared to competitors' ones. The air flow on the upper partis diverted to the rear of the appliance by a tangential flow thattouches the calibration well.
As a result, the heads of the thermocouples, which contain theconnection terminals of the compensated cables, remain at aconsiderably lower temperature reducing very much the compensationerrors produced by the heads heating.
Tehnične specifikacije
Technical specification PYROS 375
Operating range
(t amb. 20 °C / 68 °F) amb.+10 °C ÷ +375 °C (Amb. +18 °F ÷ 707 °F)
Mean heating time
(stabilization included) From 30 °C / 86 °F to 375 °C / 707 °F 20 min
Mean cooling time
(stabilization included) From 375 °C / 707 °F to 100 °C / 212 °F 40 min
Axial uniformity ±0,10 °C at 50 °C ±0.1 °F at 122 °F
±0,02 °C at 50 °C ±0.03 °F at 122 °F
±0,05 °C at 150 °C (±32.09 °F at 302 °F)
±0,15 °C at 450 °C (±32.27 °F at 842 °F)
±0,15 °C at 375 °C (±0.27 °F at 707 °F)
±0,35 °C at 650 °C (±0.6 °F at 1202 °F)
Radial uniformity (at 40 mm) ±0,10 °C at 50 °C (±0.1 °F at 122 °F)
±0,05 °C at 50 °C (±0.09 °F at 122 °F)
±0,15 °C at 150 °C (±0.27 °F at 302 °F)
±0,10 °C at 150 °C (±0.1 °F at 302 °F)
±0,20 °C at 375 °C (±0.3 °F at 707 °F)
±0,15 °C at 375 °C (±0.27 °F at 707 °F)
Hole depth 150 mm (5.9 in.)
Hole diameter 26 mm (1 in.)
Display accuracy (±1 digit) ±0,25 °C at 150 °C (±0.45 °F at 302 °F)
±0,25 °C at 150 °C (±0.45 °F at 302 °F)
±0,35 °C at 375 °C (±0.6 °F at 707 °F)
±0,35 °C at 375 °C (±0.6 °F at 707 °F)
Stability ±0,15 °C - (±0.27 °F)
±0,10 °C - (±0.1 °F)
Display Resolution 0,1 °C - (0.1 °F)
Units of measure °C / °F
Over temperature cut out Yes
Switch test 5 VAC DC PC
Interface RS 232
Display LED double line IN / OUT RS232
Temperature ramps Yes
Calibration From 1 to 10 calibration points
Calibrator size 130 x 260 x 280 mm (5.1 x 10 x 11 in.)
Calibrator weight 5,4 kg - (11 lb)
Power supply Automatic switchable voltage supply
115/230 VAC ±10% - 50/60 Hz
Electric power 600 W